Dionex ICS-5000
+
Installation Instructions
36
Doc. 065447-02 12/14
3.5
Setting Up the DC for Analytical IC Applications
is an example of the interior of the DC after the components for
two analytical IC systems have been installed.
3.5.1
Installing the CD or ED
NOTE
Before installing an ED (P/N 072042), follow the instructions
in
to install the ED amperometry cell.
1. Verify that the main power switch on the rear panel of the DC is off.
2. Pull out the DC upper compartment tray until it stops.
3. Install the analytical CD (P/N 079829) and/or the ED in the bottom half of
the DC upper compartment (see
•
For a single-system configuration, it does not matter on which side of
the compartment the detector is installed.
•
For a dual-system configuration, install each detector above the
corresponding injection valve.
Figure 3-7. Example DC Interior View: Dual-Analytical System
Automation
Manager
Components
Columns and
Injection Valves
Electrochemical
Detector and Cell
Conductivity
Detector and
Suppressor
Upper Compartment
Lower Compartment
1
2
3
4
1
2
3
4