3 • System Setup
Doc. 065447-02 12/14
65
1. Rinse the eluent reservoir with deionized water and then fill it with
degassed deionized water. This water will be used to:
•
Flush the system
•
Condition and hydrate consumable devices (EGC, CR-TC,
suppressor, CRD)
•
Generate eluent (in RFIC-EG systems)
For non-RFIC-EG systems:
You can prepare the eluent for the
application now or wait until after completing the system installation and
the IQ. For eluent preparation instructions, refer to the manual for the
column that will be installed. Column manuals are provided on the
Thermo Scientific Reference Library DVD (P/N 053891).
2. Place the reservoir in the Dionex ICS-5000
+
Eluent Organizer (EO):
•
EO with four 2-liter reservoirs (P/N 072058)
•
EO with two 2-liter reservoirs (P/N 072057)
•
EO without reservoirs (P/N 072059)
If an EO is not available, place the reservoir in appropriate secondary
containment.
Verwenden Sie den Plastikbehälter zum Offline Vakuum-Entgasen von
Eluenten. Die Behälter sind dafür nicht ausgelegt.
Always use ASTM filtered, Type I (or better) (18 megohm-cm) deionized
water that meets the specifications in
to rinse
reservoirs, fill reservoirs used for eluent generation or seal washing, or
to manually prepare eluent.