3 • System Setup
Doc. 065447-02 12/14
55
•
Click
Configure
next to
Valve
and verify that the IC Cube injection valve
is assigned to the correct instrument (or timebase) (see
For
Controlled by
, verify that
AS
is selected. This setting assigns control
of the valve to the autosampler. Use this setting for both the Dionex AS-
AP and Dionex AS-DV. If an autosampler is not installed, select
DC
in
the
Controlled by
list.
Figure 3-24. IC Cube Valve Configuration Example