Installation Instructions
130
Doc. 065447-02 12/14
B.1.2
Installing a Disposable Working Electrode
For installation instructions for disposable working electrodes, refer to the
installation guide shipped with the electrodes (see below) or to
Product
Manual for Disposable Electrodes
(Document No. 065040), which is
provided on the Thermo Scientific Reference Library DVD (P/N 053891).
•
Disposable Silver Electrode Installation Guide for ED
(Document
No. 065137)
•
Disposable Platinum Electrode Installation Guide for ED
(Document
No. 065139)
•
Disposable Gold Electrode Installation Guide
(Document No. 065191)
After installing a disposable working electrode, install the reference electrode:
•
To install a pH-Ag/AgCl reference electrode, go on to
•
To install a PdH reference electrode, go on to
B.1.3
Installing a Conventional Working Electrode
Notes on Polishing Conventional (Nondisposable) Working Electrodes
•
Do not polish new conventional working electrodes before
installation.
After an electrode has been used for a period of time, a layer of contamination
may build up. When this occurs, the electrode must be polished. For details,
refer to
Polishing Gold Working Electrodes
(Document No. 031154). The
manual is provided on the Thermo Scientific Reference Library DVD
(P/N 053891).
When you install a disposable electrode, be sure to install the correct
gasket for your system:
•
For a capillary system, use a 0.001 in PTFE gasket (P/N 072117,
Pkg. of 2).
•
For an analytical system with a gold, silver, or platinum
disposable electrode, use a 0.002 in PTFE gasket (P/N 060141,
Pkg. of 4).
•
For an analytical system with a carbon disposable electrode, use
a 0.001 in Ultem
™
gasket (P/N 069339).