Installation Instructions
94
Doc. 065447-02 12/14
2. Route tubing inside the DC through the tubing trays below the IC Cube
a.
Press the sample in and sample out lines from the injection valve into
the tubing tray on the front of the IC Cube and then route the lines
through the DC left-side slot to the autosampler.
b. If you are installing a dual system:
•
Press the sample in and sample out lines from both valves into the
tubing trays and then route all four lines through the DC left-side
slot to the autosampler (see
•
Press the
Eluent In
line from the EG degas cartridge of system #1
into the tubing trays below the IC Cube of system #2 and then
route the line out the DC right-side slot to the EG.
Figure 4-18. IC Cube Tubing Trays (System #2 of Dual System Shown)
Figure 4-19. Completed IC Cube Plumbing for a Dual Conductivity
Detection System
Tubing Tray