B • Installing DC Options
Doc. 065447-02 12/14
139
B.1.5
Installing the PdH Reference Electrode in a Capillary IC
System
1. To avoid any hydraulic pressure buildup when inserting the reference
electrode into the cell, make sure that fitting plugs are not installed on the
cell inlet and outlet fittings.
2. Using tweezers, grasp the edge of the PdH reference electrode gasket
(P/N 072214) (see
). To avoid deforming the gasket cutout, do
not place the tweezer tips on the cutout.
3. Orient the gasket so that the notched edge of the gasket aligns with the
corresponding notch in the reference electrode well. Then, press the
gasket into the well (see
Capillary
IC
Figure B-15. Installing the PdH Reference Electrode Gasket
Figure B-16. Reference Electrode Well with
PdH Reference Electrode Gasket Installed
PdH Reference
Electrode Gasket
Cutout
Notch
PdH Reference
Electrode Gasket
Notch in
Well