4 • System Plumbing
Doc. 065447-02 12/14
109
8. In Chromeleon:
a.
Verify that the suppressor and Dionex EGC are turned off.
b.
Gradient pump only:
Enter 100% for the reservoir containing the
deionized water.
c.
Set the pump flow to the rate specified in the Quality Assurance
Report (QAR) shipped with your separator column.
9. Allow the pump to run for at least 10 minutes. Turn off the pump flow.
10. Disconnect the hydration tubing assembly from the
ELUENT OUT
and
REGEN IN
ports on the Dionex CR-TC.
Completing the Dionex CR-TC Plumbing
Refer to
and the following steps to complete the Dionex CR-TC
plumbing.
To avoid damaging the suppressor, always turn off the suppressor
before hydrating the Dionex CR-TC. The pump flow is on during
conditioning; however, no flow reaches the suppressor.
Figure 4-30. Dionex CR-TC Plumbing Diagram: Analytical Flow Version
Installation Completed
To RFIC Eluent
Degasser
ELUENT IN Port
Coupler
To
OUTLET Port
EGC
WASTE, GAS
SEPARATOR
TO CR
-
TC
REGEN OUT
(blue label)
CR-TC
To CR-TC
Eluent In
(red label)
To CR-TC
Eluent Out
(yellow label)
E
L
U
E
N
T
O
U
T
R
E
G
E
N
I
N
To CR-TC
Regen In
(orange label)
TO SRS/AES
REGEN OUT
+