Dionex ICS-5000
+
Installation Instructions
60
Doc. 065447-02 12/14
3. Place the open end of each hose into a waste container or appropriate
drain.
Connecting a Rigid Drain Line to a Low-Temperature DC
The low-temperature DC includes parts for assembling a rigid drain line that
can be routed from the drain port on the DC rear panel to either the rear, side,
or front edge of the lab bench. The standard corrugated drain hose then
connects to the rigid drain line and is directed to the waste container or drain.
This design prevents the drain line from being accidentally elevated or
crimped as it is routed to the drain. A vent to help prevent vapor lock and
improve liquid flow to the drain is also included.
1. Locate the drain line parts in the low-temperature DC Ship Kit
(P/N 22171-62000) (see
):
For correct drainage, make sure drain hoses lay flat on the bench and
are not bent, pinched, or elevated at any point. Do not allow the ends of
the drain hoses to be submerged in waste liquid.
Part
Part Number
Quantity
Corrugated drain hose
055075
1
PTFE tubing, 1.3 cm (0.5 in) OD,
6.5 cm (2.5 in) long
22171-30001
2
PTFE
tubing,
1.3 cm (0.5 in)
OD,
33.0 cm
(13.0 in)
long
22171-30002
3
Fitting
tee,
push
‐
in,
RoHS,
1.3 cm
(0.5 in)
00101-06-00010
1
Fitting
elbow,
push
‐
in,
RoHS,
1.3 cm
(0.5 in)
00101-05-00008
3
Fitting
union,
push
‐
in,
RoHS,
1.3 cm
(0.5 in)
00101-01-00049
1
Nylon
vent
plug,
RoHS,
1.3 cm (0.5 in)
00201-99-00406
1
Table 3-3. Parts for the DC Drain Line