Dionex ICS-5000
+
Installation Instructions
90
Doc. 065447-02 12/14
4.1.12 Completing the Plumbing for Conductivity Detection
1. Connect the CES
ELUENT IN
line to the column
ELUENT OUT
2. If a CRD is installed, connect the CES
ELUENT OUT
line to the CRD
ELUENT IN
port and connect the CRD
ELUENT OUT
line to the CD
IN
port.
Capillary
IC
Figure 4-16. Flow Schematic for Equilibration and Operation of a
Conductivity Detection System with a CRD Installed
EGC
(capillary)
OUT
IN
E
L
U
E
N
T
O
U
T
R
E
G
E
N
IN
CR-TC
ELUENT OUT
ELUENT IN
REGEN OUT
REGEN IN
5
6
3
CRD
CES
2
SAMPLE OUT
REGEN IN
ELUENT OUT
ELUENT IN
SAMPLE IN
ELUENT OUT
ELUENT IN
ELUENT OUT
ELUENT IN
ELUENT OUT
ELUENT IN
EG
DEGAS
ELUENT IN
ELUENT OUT
VENT
5000 psi
Conductivity Detector
(Capillary)
CELL IN
CELL OUT
To waste
From pump
To gas separator
waste tube
From
autosampler
Through
tubing chase
to rear panel