3. Install an end-line filter (P/N 045987) on each reservoir’s eluent line (if
they are not already installed). The filters are in the pump ship kit:
4. Thoroughly rinse the end-line filters with deionized water.
5. Install the reservoir cap, making sure the end of the line extends to the
bottom of the reservoir, and that the filter is submerged in liquid. This
prevents air from being drawn through the eluent lines. Hand-tighten the
cap.
NOTE
A Dionex High-Pressure Inline Filter (P/N 044105)
can be used to remove microparticulate matter from
eluent. Connect the inline filter between the pump out-
let and the eluent inlet port on the injection valve. For
details, see the instructions provided with the inline
filter.
6. Locate the 10 mL syringe (P/N 079803) in the pump ship kit and a luer
adapter (P/N 024305) in the DC Ship Kit (standard DC, P/N 072011; low-
temperature DC, P/N 22171-62000).
7. Using the syringe and luer adapter, draw water through each reservoir
ELUENT BOTTLE OUT
line to flush the line.
8. Connect the liquid lines from the DP/SP to the
ELUENT BOTTLE OUT
line
of each reservoir. If an eluent line is not used, retain the cap on the end of
the line.