6. If you are installing system 1:
a.
Orient the Dionex CR-TC with the
ELUENT IN
port (yellow label)
toward the front of the EG and push the Dionex CR-TC firmly onto
the ball studs on the left side of the EG.
b. Route the Dionex EGC and CR-TC electrical cables to the right side
of the EG tray.
7. If you are installing system 2:
Orient the Dionex CR-TC with the
ELUENT IN
port (yellow label) toward
the back of the EG and push the Dionex CR-TC firmly onto the ball studs
on the right side of the EG.
8. Connect the CR-TC and EGC electrical cables:
a.
Align the three pins inside the Dionex CR-TC cable connector with
the holes in the
CR-TC 1
(or
CR-TC 2
) bulkhead connector. Push the
cable connector firmly onto the connector (see
). Twist the
locking ring on the cable connector fingertight to secure it.
b. Align the four pins inside the Dionex EGC cable connector with the
holes in the
EGC 1
(or
EGC 2
) bulkhead connector. Push the cable
connector firmly onto the bulkhead connector (see
). Twist
the ring on the cable connector fingertight to secure it.