Dionex ICS-5000
+
Installation Instructions
68
Doc. 065447-02 12/14
2. Place the cap on the reservoir and tighten fingertight.
3. Insert the reservoir into the holder on the component panel.
4. Check that the tubing is engaged in the peristaltic pump (see
If it is not, lift the lever on the peristaltic pump to the right, insert the
tubing neatly between the lever and the rotor, and release the lever to
secure the tubing.
3.13.2 Activating the Piston Seal Wash Function
1. Open the Chromeleon 7 ePanel Set or the Chromeleon 6.8 panel tabset.
2. Press the
F8
key to open the Command window (Chromeleon 7) or the
Commands dialog box (Chromeleon 6.8).
3. Select the pump name.
4. For Chromeleon 7, click the
Properties
tab in the right pane of the
window. For Chromeleon 6.8, scroll through the list of commands and
properties under the pump name.
5. Select the
RearSealWashSystem
property and select the
Interval
option.
6. For Chromeleon 6.8, click the
Execute
button.
Figure 3-34. Peristaltic (Piston Seal Wash) Pump
Lever
Outlet Tubing
from Seal
Wash
Reservoir
Tubing to
Secondary
Pump Head
Rotor