Dionex ICS-5000
+
Installation Instructions
128
Doc. 065447-02 12/14
2. Loosen the yoke knob that holds the spacer block to the cell body by
unscrewing it two to three turns.
3. Squeeze the tabs on the sides of the yoke block and pull the knob with
yoke block off the spacer block.
NOTE
Handle the inside surfaces of the cell carefully to pre-
vent scratches, which may subsequently cause leaks.
4. Slide the spacer block off the alignment pins on the cell body (see
). The spacer block is for use only with disposable working
electrodes. If you will be installing a conventional working electrode, you
can remove the spacer block and save it for possible future use.
Figure B-2. Disassembling the ED Cell
Spacer Block
Cell Body