Dionex ICS-5000
+
Installation Instructions
34
Doc. 065447-02 12/14
3.4.2
Installing the EG Degas, CRD, or CES Cartridge
1. To install an EG degas cartridge:
a.
Remove the cap from the regenerant port on the rear of the cartridge.
Slide the cartridge into the right-most slot of the Dionex IC Cube (see
). Push the cartridge firmly into the slot and tighten the
thumbscrews fingertight only.
b. Connect the green PEEK vent line (P/N 088271) supplied with the
cartridge to the
VENT
port on the cartridge. Route the vent line
through the tubing chase to the rear panel.
2. To install a CRD or CES cartridge:
a.
Loosen the thumbscrews on the CRD or CES bypass cartridge and
pull the bypass cartridge out of the Dionex IC Cube. Save the bypass
cartridge.
a.
Remove the caps from the regenerant ports on the rear of the CRD or
CES cartridge. Slide the cartridge into its slot in the Dionex IC Cube
(see
). Push the cartridge firmly into the slot and tighten the
thumbscrews fingertight.
NOTE
Install the saved regenerant port caps onto the CRD
and CES cartridge ports. Save the bypass cartridges.
The bypass cartridges must be reinstalled if the
Dionex IC Cube is used for an application that does
not require a CRD or suppressor.
3.4.3
Installing the Detector in the DC
Installing the CD or ED
NOTE
Before installing an ED (P/N 072042), follow the instructions
in
to install the ED amperometry cell.
1. Verify that the main power switch on the rear panel of the DC is off.
Capillary
IC
Capillary
IC
Capillary
IC