Dionex ICS-5000
+
Installation Instructions
108
Doc. 065447-02 12/14
10. Reconnect the Dionex EGC electrical cable.
Hydrating the Dionex CR-TC
1. Remove the plugs from the Dionex CR-TC ports.
2. Locate the CR-TC Hydration Tubing Kit (P/N 063487) provided in the
EG Ship Kit (P/N 072047).
3. Connect the hydration tubing as shown in
.
4. Connect the Dionex EGC outlet line to the CR-TC
ELUENT IN
port.
5. Disconnect the 1.58 mm (0.062 in) ID ETFE (ethylene
tetrafluoroethylene) tubing labeled
WASTE, GAS SEPARATOR / SRS/AES
REGEN OUT
from the system tubing package (described on
6. Connect this tubing to the CR-TC
REGEN OUT
port.
7. Route the
WASTE, GAS SEPARATOR
line through the slot on the EG and
connect it to the gas separator waste tube. For instructions, see
.
Figure 4-29. Dionex CR-TC Plumbing for Hydration: Analytical Flow
Version
From EGC Outlet Port
CR-TC Hydration Kit
Tubing Assembly
(P/N 063487)
(Temporary connection)
To Waste
E
L
U
E
N
T
O
U
T
R
E
G
E
N
I
N
TO CR
-
TC
REGEN OUT
(blue label)
To CR-TC
Eluent In
(red label)
Eluent Out
(yellow label)
Regen In
(orange label)
Dionex CR-TC Hydration:
Pump at QAR rate for 10 min