4 • System Plumbing
Doc. 065447-02 12/14
95
4.2
Plumbing an Analytical IC System
4.2.1
Tubing and Fittings
One tubing package with precut and labeled tubing (either microbore or
standard bore) is included for each system configured in the ICS-5000
+
. The tubing is used for plumbing the columns, detector cell,
suppressor, and system waste. Tubing and fittings for plumbing an injection
valve waste line are also included in each package.
Before connecting each piece of tubing, remove the unions that connect the
tubing in the tubing package. Use tubing with white labels for system 1 (left
side) and tubing with blue labels for system 2 (right side).
Tubing and fittings for other connections (for example, the pump and Dionex
EGC outlet lines) are provided in the module ship kits.
•
0.25 mm (0.010 in) ID black PEEK tubing (P/N 052306) is provided for
standard bore systems
•
0.125 mm (0.005 in) ID red PEEK tubing (P/N 052310) is provided for
microbore systems
Plumbing
Configuration
System
Number
Color of Tubing
Labels
Part
Number
Microbore
1 White
062035
2
Blue
062036
Standard Bore
1
White AAA-062031
2
Blue
062062
Table 4-1. System Tubing Packages