4. Note which rear panel connector each valve is connected to. This
information is required to control the valves in Chromeleon.
B.2.2
Installing a Low-Pressure Valve Inside the DC
1. Insert the two tabs on the valve mounting plate into available slots in the
lower compartment of the DC, above the columns.
2. While holding the valve mounting plate firmly in place, tighten the
captive thumbscrew.
3. Route the valve cable to the back of the DC, through the slot on the side
of the module. Connect the cable to one of the low-pressure valve
connectors on the DC rear panel (see
Figure B-19. DC Rear Panel: External Low-Pressure Valve Connections
External Low-
Pressure Valve
Connectors
Location of Valve
Mounting Slots
1
2
1
2