Dionex ICS-5000
+
Installation Instructions
132
Doc. 065447-02 12/14
3. Orient the working electrode block as shown in
it onto the alignment pins.
4. Install the yoke knob on the alignment pins and fingertighten the knob by
turning it a full 360 degrees.
NOTE
It is not possible to overtighten the yoke knob. After
the knob clicks into place, it cannot be tightened any
further. The yoke knob ensures that constant pressure
is applied to the cell.
5. Install the reference electrode:
•
If you are installing a pH-Ag/AgCl reference electrode, go on to
.
•
If you are installing a PdH reference electrode, go on to
.
Figure B-6. ED Cell Components
Conventional
Working
Electrode Block
Cell
Body
Cell
Gasket
Yoke Knob and
Block