Operation
Alarms Menu
3-60
Model 43i Trace Level-Enhanced Instruction Manual
Thermo Fisher Scientific
the minimum or maximum limit, an alarm is activated. The word
“ALARM” appears in the Run screen and in the Main Menu.
In the Main Menu, choose Alarms >
Pressure
.
PRESSURE:
ACTUAL 750.0 mmHg
>MIN 400.0 mmHg
MAX 1000.0 mmHg
RANGE AVG DIAGS ALARM
The Minimum Pressure alarm limit screen is used to change the minimum
pressure alarm limit. The minimum and maximum pressure screens
function the same way.
In the Main Menu, choose Alarms > Pressure >
Min
or
Max
.
PRESSURE:
ACTUAL MIN: 400.0 mmHg
SET MIN TO: 500.0 mmHg?
©ª
INC/DEC
SAVE VALUE
RANGE AVG DIAGS ALARM
The Sample Flow screen displays the current sample flow reading and sets
the minimum and maximum alarm limits. Acceptable alarm limits range
from 0 to 2.5 LPM. If the sample flow reading goes beyond either the
minimum or maximum limit, an alarm is activated. The word “ALARM”
appears in the Run screen and in the Main Menu.
In the Main Menu, choose Alarms >
Sample Flow
.
SAMPLE FLOW:
ACTUAL 0.550 LPM
>MIN 0.000 LPM
MAX 1.000 LPM
RANGE AVG DIAGS ALARM
Min and Max Pressure
Limits
Sample Flow
Содержание 43i
Страница 4: ......
Страница 18: ...Contents Where to Get Help xviii Model 43i Trace Level Enhanced Instruction Manual Thermo Fisher Scientific ...
Страница 28: ...Introduction Specifications 1 6 Model 43i Trace Level Enhanced Instruction Manual Thermo Fisher Scientific ...
Страница 38: ...Installation Startup 2 10 Model 43i Trace Level Enhanced Instruction Manual Thermo Fisher Scientific ...
Страница 222: ...System Description I O Components 8 14 Model 43i Trace Level Enhanced Instruction Manual Thermo Fisher Scientific ...
Страница 238: ......
Страница 312: ......