![Thermo Scientific 43i Скачать руководство пользователя страница 59](http://html.mh-extra.com/html/thermo-scientific/43i/43i_instruction-manual_1105147059.webp)
Operation
Calibration Menu
Thermo Fisher Scientific
Model 43i Trace Level-Enhanced Instruction Manual
3-21
The Zero Duration Minutes screen defines how long zero air is sampled by
the instrument. The span and purge duration screens look and functions
the same way as the zero duration screen, and are used to set how long the
span gas and sample gas are sampled by the instrument. Durations between
0 and 60 minutes are acceptable. Each time a zero/span check occurs the
zero check is done first, followed by the span check. To perform just a zero
check, set the span and purge duration screens to 0 (off). The same applies
to perform just a span or purge check.
In the Main Menu, choose Calibration > Zero/Span Check >
Zero, Span
or
Purge Duration Min
.
ZERO DURATION:
CURRENTLY: 30 MIN
SET TO: 31 MIN ?
§¨
MOVE CURSOR
©ª
CHANGE VALUE
SAVE
RANGE AVG DIAGS ALARM
The Zero/Span Averaging Time screen allows the user to adjust the
zero/span averaging time. The zero/span averaging time is used by the
analyzer only when performing an automatic zero, or span check, or
calibration. The analyzer’s averaging time is used for all other functions.
The following averaging times are available: 1, 2, 5, 10, 20, 30, 60, 90,
120, 180, 240, and 300 seconds.
In the Main Menu, choose Calibration > Zero/Span Check >
Zero
/
Span
Avg Sec
.
ZERO/SPAN AVERAGING TIME:
CURRENTLY: 60 SEC
SET TO: 90 SEC ?
©ª
CHANGE VALUE
SAVE VALUE
RANGE AVG DIAGS ALARM
The Zero/Span Ratio screen is used to adjust the ratio of zeros to spans. For
example, if this value is set to 1, a span check will follow every zero check.
If this value is set to 3, there will be two zero checks between each
zero/span check. This value may be set from 1 to 10, with 1 as default.
Zero/Span/Purge
Duration Minutes
Zero/Span Averaging
Time
Zero/Span Ratio
Содержание 43i
Страница 4: ......
Страница 18: ...Contents Where to Get Help xviii Model 43i Trace Level Enhanced Instruction Manual Thermo Fisher Scientific ...
Страница 28: ...Introduction Specifications 1 6 Model 43i Trace Level Enhanced Instruction Manual Thermo Fisher Scientific ...
Страница 38: ...Installation Startup 2 10 Model 43i Trace Level Enhanced Instruction Manual Thermo Fisher Scientific ...
Страница 222: ...System Description I O Components 8 14 Model 43i Trace Level Enhanced Instruction Manual Thermo Fisher Scientific ...
Страница 238: ......
Страница 312: ......