![Thermo Scientific 43i Скачать руководство пользователя страница 251](http://html.mh-extra.com/html/thermo-scientific/43i/43i_instruction-manual_1105147251.webp)
C-Link Protocol Commands
Alarms
Thermo Fisher Scientific
Model 43
i
Trace Level-Enhanced Instruction Manual
B-11
alarm chamber temp min
alarm chamber temp max
These commands report the chamber temperature alarm minimum and
maximum value current settings. The example that follows reports that the
chamber temperature alarm minimum value is 35.0
°
C.
Send:
alarm
chamber
temp
min
Receive:
alarm chamber temp min 35.0 deg C
set alarm chamber temp min
value
set alarm chamber temp max
value
These commands set the chamber temperature alarm minimum and
maximum values to
value
, where
value
is a floating-point number
representing chamber temperature alarm limits in degrees C. The example
that follows sets the chamber temperature alarm maximum value to 55.0
°
C.
Send:
set alarm chamber temp max 55.0
Receive:
set alarm chamber temp max 55.0 ok
alarm conc so2 min
alarm conc so2 max
These commands report the SO
2
concentration alarm minimum and
maximum values current setting. The example that follows reports that the
SO
2
concentration minimum is 5.2 ppb.
Send:
alarm conc so2 min
Receive:
alarm conc so2 min 5.2 ppb
set alarm conc so2 min
value
set alarm conc so2 max
value
These commands set the SO
2
concentration alarm minimum and
maximum values to
value,
where
value
is a floating-point representation of
the concentration alarm limits. Values must be in the ppb or
μg/m
3
. Refer
to “Entering Units in PPB” in this chapter. The example that follows sets
the SO
2
concentration alarm maximum value to 80 ppm.
Send:
set alarm conc so2 max 80000
Receive:
set alarm conc so2 max 80000 ok
Alarms
Содержание 43i
Страница 4: ......
Страница 18: ...Contents Where to Get Help xviii Model 43i Trace Level Enhanced Instruction Manual Thermo Fisher Scientific ...
Страница 28: ...Introduction Specifications 1 6 Model 43i Trace Level Enhanced Instruction Manual Thermo Fisher Scientific ...
Страница 38: ...Installation Startup 2 10 Model 43i Trace Level Enhanced Instruction Manual Thermo Fisher Scientific ...
Страница 222: ...System Description I O Components 8 14 Model 43i Trace Level Enhanced Instruction Manual Thermo Fisher Scientific ...
Страница 238: ......
Страница 312: ......