![Thermo Scientific 43i Скачать руководство пользователя страница 270](http://html.mh-extra.com/html/thermo-scientific/43i/43i_instruction-manual_1105147270.webp)
C-Link Protocol Commands
Measurement Configuration
B-30
Model 43
i
Trace Level-Enhanced Instruction Manual
Thermo Fisher Scientific
Send:
gas
mode
Receive: gas
mode
sample
set sample
set sample gas
This command sets the zero/span valves to the sample mode. The example
that follows sets the instrument to sample mode, that is, the instrument is
reading the sample gas.
Send:
set
sample
Receive: set
sample
ok
set zero
set zero gas
This command sets the zero/span valves to the zero mode. The example
that follows sets the instrument to zero mode that is, the instrument is
reading the sample gas.
Send:
set
zero
Receive: set
zero
ok
set span
set span gas
This command sets the zero/span valves to the span mode. The example
that follows sets the instrument to span mode that is, the instrument is
sampling span gas.
Send:
set
span
Receive: set
span
ok
gas unit
This command reports the current gas units (ppb or
μ
g/m
3
). The example
reports that the gas unit is set to ppm.
Send:
gas
unit
Receive: gas
unit
ppm
set gas
unit
unit
= | ppm | mg/m
3
|
This command sets the gas units to ppm or mg/m
3
. The example that
follows sets the gas units to mg/m
3
.
Send:
set
gas
unit
mg/m3
Receive:
set gas unit mg/m3 ok
Содержание 43i
Страница 4: ......
Страница 18: ...Contents Where to Get Help xviii Model 43i Trace Level Enhanced Instruction Manual Thermo Fisher Scientific ...
Страница 28: ...Introduction Specifications 1 6 Model 43i Trace Level Enhanced Instruction Manual Thermo Fisher Scientific ...
Страница 38: ...Installation Startup 2 10 Model 43i Trace Level Enhanced Instruction Manual Thermo Fisher Scientific ...
Страница 222: ...System Description I O Components 8 14 Model 43i Trace Level Enhanced Instruction Manual Thermo Fisher Scientific ...
Страница 238: ......
Страница 312: ......