![Thermo Scientific 43i Скачать руководство пользователя страница 278](http://html.mh-extra.com/html/thermo-scientific/43i/43i_instruction-manual_1105147278.webp)
C-Link Protocol Commands
I/O Configuration
B-38
Model 43
i
Trace Level-Enhanced Instruction Manual
Thermo Fisher Scientific
set instrument id
value
This command sets the instrument id to
value,
where
value
is a decimal
number between 0 and 127 inclusive.
Note
Sending this command via RS-232 or RS-485 will require the host to
use the new id for subsequent commands.
▲
Send:
set
instrument
id
12
Receive:
set instrument id 12 ok
mode
This command reports what operating mode the instrument is in: local,
service, or remote. The example that follows shows that the instrument is
in the remote mode.
Send:
mode
Receive: mode
remote
set mode local
set mode remote
These commands set the instrument to local or remote mode. The example
that follows sets the instrument to the local mode.
Send:
set
mode
local
Receive:
set mode local ok
program no
This command reports the analyzer’s model information and program
version number, which will be dependant on the current version.
Send:
program
no
Receive:
program no iSeries 43i 01.01.10.003
set layout ack
This command disables the stale layout/layout change indicator (‘*’) that is
attached to each response if the layout has changed.
Send:
set
layout
ack
Receive:
set layout ack ok
analog iout range
channel
This command reports the analog current output range setting for
channels
,
where
channel
must be between 1 and 6, inclusive. The example that
follows reports current output channel 4 to the 4-20 mA range, according
I/O Configuration
Содержание 43i
Страница 4: ......
Страница 18: ...Contents Where to Get Help xviii Model 43i Trace Level Enhanced Instruction Manual Thermo Fisher Scientific ...
Страница 28: ...Introduction Specifications 1 6 Model 43i Trace Level Enhanced Instruction Manual Thermo Fisher Scientific ...
Страница 38: ...Installation Startup 2 10 Model 43i Trace Level Enhanced Instruction Manual Thermo Fisher Scientific ...
Страница 222: ...System Description I O Components 8 14 Model 43i Trace Level Enhanced Instruction Manual Thermo Fisher Scientific ...
Страница 238: ......
Страница 312: ......