Operation
Service Menu
Thermo Fisher Scientific
Model 43i Trace Level-Enhanced Instruction Manual
3-79
The Set Defaults screen allows the user to reset the configuration values to
factory defaults.
In the Main Menu, choose Service > Perm Oven Settings > Factory Cal
Oven Therm >
Set Defaults
.
RESTORE DEFAULT SETTINGS:
FOR PERM OVEN HTR THERM
RESTORE
ARE YOU SURE YOU WANT TO?
x
PRESS
¨
TO CONFIRM RESTORE
RANGE AVG DIAGS ALARM
The Dilution Ratio screen allows the user to view and set the dilution ratio.
Acceptable values are 1–500: 1. The default is 1:1. When this value is set,
the dilution ratio is applied to all concentration measurements. This screen
is only accessible if the dilution ratio option is installed.
In the Main Menu, choose Service >
Dilution Ratio
.
DILUTION RATIO:
CURRENTLY: 001.0 :1
SET TO: 00
2
.0 :1 ?
§¨
MOVE CURSOR
©ª
CHANGE VALUE
SAVE
The Display Pixel Test is used to test the LCD display. The display pixel
test screen is visible only when the instrument is in service mode. For more
information on the service mode, see “Service Mode” earlier in the chapter.
In the Main Menu, choose Service >
Display Pixel Test
.
Set defaults
Dilution Ratio
Display Pixel Test
Содержание 43i
Страница 4: ......
Страница 18: ...Contents Where to Get Help xviii Model 43i Trace Level Enhanced Instruction Manual Thermo Fisher Scientific ...
Страница 28: ...Introduction Specifications 1 6 Model 43i Trace Level Enhanced Instruction Manual Thermo Fisher Scientific ...
Страница 38: ...Installation Startup 2 10 Model 43i Trace Level Enhanced Instruction Manual Thermo Fisher Scientific ...
Страница 222: ...System Description I O Components 8 14 Model 43i Trace Level Enhanced Instruction Manual Thermo Fisher Scientific ...
Страница 238: ......
Страница 312: ......