![Thermo Scientific 43i Скачать руководство пользователя страница 248](http://html.mh-extra.com/html/thermo-scientific/43i/43i_instruction-manual_1105147248.webp)
C-Link Protocol Commands
Measurements
B-8
Model 43
i
Trace Level-Enhanced Instruction Manual
Thermo Fisher Scientific
Send:
avg
time
Receive:
avg time 11:300 sec
set avg time
selection
set high avg time
selection
set low avg time
selection
These commands set the averaging time, high and low averaging times,
according to
Table B–2
. The example that follows sets the low range
averaging time to 120 seconds.
Send:
set
low
avg
time
8
Receive:
set low avg time 8 ok
Table B–2.
Averaging Times
Selection
Averaging Time (seconds)
0 1
seconds
1 2
2 5
3 10
4 20
5 30
6 60
7 90
8 120
9 180
10 240
11 300
so2
high so2
low so2
These commands report the measured SO
2
concentration when operating
in single range, or high and low SO
2
when operating in dual or auto range
mode. The example that follows shows that the SO
2
concentration is 40
ppm.
Send:
so2
Receive:
so2 0040E+0 ppm
Содержание 43i
Страница 4: ......
Страница 18: ...Contents Where to Get Help xviii Model 43i Trace Level Enhanced Instruction Manual Thermo Fisher Scientific ...
Страница 28: ...Introduction Specifications 1 6 Model 43i Trace Level Enhanced Instruction Manual Thermo Fisher Scientific ...
Страница 38: ...Installation Startup 2 10 Model 43i Trace Level Enhanced Instruction Manual Thermo Fisher Scientific ...
Страница 222: ...System Description I O Components 8 14 Model 43i Trace Level Enhanced Instruction Manual Thermo Fisher Scientific ...
Страница 238: ......
Страница 312: ......