![Thermo Scientific 43i Скачать руководство пользователя страница 171](http://html.mh-extra.com/html/thermo-scientific/43i/43i_instruction-manual_1105147171.webp)
Servicing
Removing the Measurement Bench and Lowering the Partition Panel
Thermo Fisher Scientific
Model 43i Trace Level-Enhanced Instruction Manual
7-7
Figure 7–3.
Removing the Measurement Bench and Lowering the Partition
Panel
Equipment Required:
Philips screwdriver
Equipment Damage
Some internal components can be damaged by small
amounts of static electricity. A properly ground antistatic wrist strap must
be worn while handling any internal component.
▲
1.
Turn instrument OFF and unplug the power cord.
2.
If the instrument is mounted in a rack, remove it from the rack.
3.
Remove the cover.
4.
Disconnect the plumbing connections at the rear of the measurement
bench.
Pass Through
Connectors
Partition Panel
Loosen
Captive
Screw
Remove Screw
Remove Screw
Remove Screws
REMOVE SCREWS
Remove Hidden Screw
SLIDE BENCH OUT
LOWER PANEL