Operation
Diagnostics Menu
3-52
Model 43i Trace Level-Enhanced Instruction Manual
Thermo Fisher Scientific
The I/O Board screen (read only) is used to display the current voltage
readings on the I/O expansion board. This menu is only accessible if the
I/O expansion board is installed.
In the Main Menu, choose Diagnostics > Voltages >
I/O Board Voltages
.
I/O BOARD VOLTAGES:
3.3 SUPPLY 3.3 V
5.0 SUPPLY 5.0 V
24.0 SUPPLY 24.0 V
-3.3 SUPPLY -3.3 V
RANGE AVG DIAGS ALARM
The Temperatures screen (read only) displays the current internal
instrument temperature and chamber temperature. The internal
temperature is the air temperature measured by a sensor located on the
interface board.
In the Main Menu, choose Diagnostics >
Temperatures
.
TEMPERATURES:
>INTERNAL 34.6
o
C
CHAMBER 49.7
o
C
PERM OVEN GAS 45.0
o
C
PERM OVEN HTR 45.0
o
C
RANGE AVG DIAGS ALARM
The Pressure screen (read only) displays the current optical bench pressure.
The pressure is measured by a pressure transducer in-line with the optical
bench.
In the Main Menu, choose Diagnostics >
Pressure
.
PRESSURE: 753.0 mmHg
RANGE AVG DIAGS ALARM
I/O Board Voltages
Temperatures
Pressure
Содержание 43i
Страница 4: ......
Страница 18: ...Contents Where to Get Help xviii Model 43i Trace Level Enhanced Instruction Manual Thermo Fisher Scientific ...
Страница 28: ...Introduction Specifications 1 6 Model 43i Trace Level Enhanced Instruction Manual Thermo Fisher Scientific ...
Страница 38: ...Installation Startup 2 10 Model 43i Trace Level Enhanced Instruction Manual Thermo Fisher Scientific ...
Страница 222: ...System Description I O Components 8 14 Model 43i Trace Level Enhanced Instruction Manual Thermo Fisher Scientific ...
Страница 238: ......
Страница 312: ......