Operation
Service Menu
Thermo Fisher Scientific
Model 43i Trace Level-Enhanced Instruction Manual
3-75
The Permeation Oven Settings menu is used for setting up and calibrating
the permeation oven option. The permeation oven settings menu is visible
only when the instrument is in service mode. For more information on the
service mode, see “Service Mode” earlier in the chapter.
Note
This adjustment should only be performed by an instrument service
technician.
▲
In the Main Menu, choose Service >
Perm Oven Settings
.
PERM OVEN SETTINGS:
>CAL GAS THERMISTOR
CAL OVEN THERMISTOR
PERM OVEN SELECTION
FACTORY CAL GAS THERM
FACTORY CAL OVEN THERM
RANGE AVG DIAGS ALARM
The Calibrate Gas Thermistor submenu is used for calibrating the
permeation oven gas thermistor using a water bath or known resistance.
In the Main Menu, choose Service > Perm Oven Settings > Cal Gas
Thermistor >
Water Bath
or
Known Resistor
.
CALIBRATE GAS THERMISTOR:
>WATER BATH
KNOWN RESISTOR
RANGE AVE DIAGS ALARM
The Calibrate Gas Thermistor Water Bath screen allows the user to view
and set the permeation oven gas temperature to a known temperature
value.
Permeation Oven
Settings
Calibrate Gas Thermistor
Water Bath
Содержание 43i
Страница 4: ......
Страница 18: ...Contents Where to Get Help xviii Model 43i Trace Level Enhanced Instruction Manual Thermo Fisher Scientific ...
Страница 28: ...Introduction Specifications 1 6 Model 43i Trace Level Enhanced Instruction Manual Thermo Fisher Scientific ...
Страница 38: ...Installation Startup 2 10 Model 43i Trace Level Enhanced Instruction Manual Thermo Fisher Scientific ...
Страница 222: ...System Description I O Components 8 14 Model 43i Trace Level Enhanced Instruction Manual Thermo Fisher Scientific ...
Страница 238: ......
Страница 312: ......