Servicing
Thermo Fisher Scientific
Model 43i Trace Level-Enhanced Instruction Manual
7-23
Table 7–5.
Analog Input Channels and Rear Panel Pin Connections
Input Channel
Pin
1 1
2 2
3 3
4 5
5 6
6 7
7 9
8 10
Ground
4, 8, 11
Use the following procedure to calibrate the analog outputs if a meter
reading in the “Analog Output Testing” procedure differed by more than
one percent or after replacing the motherboard or optional I/O expansion
board.
Equipment Required:
Multimeter
1.
Connect a meter to the channel to be adjusted and set to voltage or
current as appropriate.
Figure 7–12
shows the analog output pins and
Table 7–4
identifies the associated channels.
2.
From the Main Menu, press
to scroll to Service, press
>
to scroll to Analog Out Calibration
,
and press
.
The Analog Output Cal screen displays.
Note
If Service Mode is not displayed, refer to “Accessing the Service
Mode” in this chapter, then return to the beginning of this step.
▲
3.
At the Analog Output Cal screen, press
to scroll to the voltage
channel or current channel corresponding to the rear panel terminal
pins where the meter is connected, then press
.
4.
With the cursor at Calibrate Zero, press
.
Analog Output
Calibration
Содержание 43i
Страница 4: ......
Страница 18: ...Contents Where to Get Help xviii Model 43i Trace Level Enhanced Instruction Manual Thermo Fisher Scientific ...
Страница 28: ...Introduction Specifications 1 6 Model 43i Trace Level Enhanced Instruction Manual Thermo Fisher Scientific ...
Страница 38: ...Installation Startup 2 10 Model 43i Trace Level Enhanced Instruction Manual Thermo Fisher Scientific ...
Страница 222: ...System Description I O Components 8 14 Model 43i Trace Level Enhanced Instruction Manual Thermo Fisher Scientific ...
Страница 238: ......
Страница 312: ......