Calibration
Zero/Span Check
4-12
Model 43i Trace Level-Enhanced Instruction Manual
Thermo Fisher Scientific
6.
To ensure that the span gas is being measured at atmospheric pressure
check that the flow is approximately 0.5 LPM or more, with the excess
flowing out the atmospheric bypass.
7.
Press
to monitor the span gas reading and wait for the reading to
stabilize.
8.
Record the difference between the measured SO
2
value and the actual
SO
2
span concentration used. This is the span drift since the last
adjustment. If the calibration has changed by more than ±10%, a new
calibration should be performed.