Operation
Range Menu
3-14
Model 43i Trace Level-Enhanced Instruction Manual
Thermo Fisher Scientific
ppb
μg/m
3
C1 C1
C2 C2
C3 C3
C1, C2, and C3 are custom ranges. For more information about custom
ranges, see “Set Custom Ranges” that follows.
The Set Custom Ranges menu lists three custom ranges: C1, C2, and C3.
Custom ranges are user-defined ranges. Any value between 10 ppb and
1000 ppb can be specified as a range. In the
μ
g/m
3
mode, any value
between 20
μ
g/m
3
and 2000
μ
g/m
3
can be specified as a range.
In the Main Menu, choose Range >
Set Custom Ranges
.
CUSTOM RANGES:
>CUSTOM RANGE 1 55.6
CUSTOM RANGE 2 75.0
CUSTOM RANGE 3 85.0
RANGE AVG DIAGS ALARM
The Custom Ranges screen is used to define the custom ranges.
The display shows the current custom range. The next line of the display is
used to set the range. To use the custom full-scale range, be sure to select it
(Custom range 1, 2, or 3) in the SO
2
Range screen. For more information
about selecting ranges, see “SO
2
Range” above.
In the Main Menu, choose Range > Set Custom Ranges >
Custom range 1,
2,
or
3
.
CUSTOM RANGE 1:
CURRENTLY: 55.5
SET TO: 000055.6 ?
§¨
MOVE CURSOR
©ª
CHANGE VALUE
SAVE
RANGE AVG DIAGS ALARM
Set Custom Ranges
Custom Ranges
Содержание 43i
Страница 4: ......
Страница 18: ...Contents Where to Get Help xviii Model 43i Trace Level Enhanced Instruction Manual Thermo Fisher Scientific ...
Страница 28: ...Introduction Specifications 1 6 Model 43i Trace Level Enhanced Instruction Manual Thermo Fisher Scientific ...
Страница 38: ...Installation Startup 2 10 Model 43i Trace Level Enhanced Instruction Manual Thermo Fisher Scientific ...
Страница 222: ...System Description I O Components 8 14 Model 43i Trace Level Enhanced Instruction Manual Thermo Fisher Scientific ...
Страница 238: ......
Страница 312: ......