C-Link Protocol Commands
Record Layout Definition
Thermo Fisher Scientific
Model 43
i
Trace Level-Enhanced Instruction Manual
B-43
For example:
Receive:
relay stat 0x0001 (indicates relay no 1 is set to
normally open logic, all others are normally closed)
Receive:
relay stat 0x0005 (indicates relay no 1 and 3 are
set to be normally open logic, all others are normally
closed)
set relay open
set relay open
value
set relay closed
set relay closed
value
These commands set the relay logic to normally open or closed for relay
number
value
, where
value
is the relay between 1 and 16. The example that
follows sets the relay no 1 logic to normally open.
Note
If the command is sent without an appended relay number then all
the relays are assigned the set logic of normally open/closed.
▲
Send:
set
relay
open
1
Receive:
set relay open 1 ok
The Erec, Lrec Srec layouts contain the following:
●
A format specifier for parsing ASCII responses
●
A format specifier for parsing binary responses,
In addition to these the Erec Layout contains
●
A format specifier for producing the front-panel displays.
In operation, values are read in using either the ASCII or binary format
specifiers and converted to uniform internal representations (32-bit floats
or 32-bit integers). These values are converted into text for display on the
screen using the format specifier for the front-panel display. Normally, the
specifier used to parse a particular datum from the input stream will be
strongly related to the specifier used to display it (e.g., all of the floating
point inputs will be displayed with an 'f' output specifier, and all of the
integer inputs will be displayed with a 'd' specifier).
The first line of the Layout response is the scanf-like parameter list for
parsing the fields from an ASCII ERec response. Parameters are separated
by spaces and the line is terminated by a \n (the normal line separator
character). Valid fields are:
Record Layout
Definition
Format Specifier for
ASCII Responses
Содержание 43i
Страница 4: ......
Страница 18: ...Contents Where to Get Help xviii Model 43i Trace Level Enhanced Instruction Manual Thermo Fisher Scientific ...
Страница 28: ...Introduction Specifications 1 6 Model 43i Trace Level Enhanced Instruction Manual Thermo Fisher Scientific ...
Страница 38: ...Installation Startup 2 10 Model 43i Trace Level Enhanced Instruction Manual Thermo Fisher Scientific ...
Страница 222: ...System Description I O Components 8 14 Model 43i Trace Level Enhanced Instruction Manual Thermo Fisher Scientific ...
Страница 238: ......
Страница 312: ......