Operation
Service Menu
3-68
Model 43i Trace Level-Enhanced Instruction Manual
Thermo Fisher Scientific
The Calibrate Pressure Zero screen calibrates the pressure sensor at zero
pressure.
Note
A vacuum pump must be connected to the pressure sensor before
performing the zero calibration.
▲
In the Main Menu, choose Service > Pressure Calibration >
Zero
.
CALIBRATE PRESSURE ZERO:
CURRENTLY: 0.0 mmHg
SET TO: 0.0 mmHg ?
CONNECT VACUUM PUMP AND
SAVE ZERO PRESSURE
RANGE AVG DIAGS ALARM
The Calibrate Pressure Span screen allows the user to view and set the
pressure sensor calibration span point.
Note
The plumbing going to the pressure sensor should be disconnected so
the sensor is reading ambient pressure before performing the span
calibration. The operator should use an independent barometer to measure
the ambient pressure and enter the value on this screen before
calibrating.
▲
In the Main Menu, choose Service > Pressure Calibration >
Span
.
CALIBRATE PRESSURE SPAN:
CURRENTLY: 756.0 mmHg
SET TO: 760.0 mmHg ?
§¨
MOVE CURSOR
©ª
CHANGE VALUE
SAVE
RANGE AVG DIAGS ALARM
Calibrate Pressure Zero
Calibrate Pressure Span
Содержание 43i
Страница 4: ......
Страница 18: ...Contents Where to Get Help xviii Model 43i Trace Level Enhanced Instruction Manual Thermo Fisher Scientific ...
Страница 28: ...Introduction Specifications 1 6 Model 43i Trace Level Enhanced Instruction Manual Thermo Fisher Scientific ...
Страница 38: ...Installation Startup 2 10 Model 43i Trace Level Enhanced Instruction Manual Thermo Fisher Scientific ...
Страница 222: ...System Description I O Components 8 14 Model 43i Trace Level Enhanced Instruction Manual Thermo Fisher Scientific ...
Страница 238: ......
Страница 312: ......