![Thermo Scientific 43i Скачать руководство пользователя страница 227](http://html.mh-extra.com/html/thermo-scientific/43i/43i_instruction-manual_1105147227.webp)
Optional Equipment
Internal Permeation Span Source
Thermo Fisher Scientific
Model 43i Trace Level-Enhanced Instruction Manual
9-5
There are two general approaches that can be used to calibrate the
permeation tube oven. The first is to calibrate the temperature indicator
very accurately (to better than 0.02
°
C) and to use a permeation tube whose
weight loss has been previously determined at that temperature.
Note
An error of approximately 0.1
°
C corresponds to an error of 1% in
release rate.
▲
The second approach is to note that the thermistors used to measure
temperature are interchangeable to better than
±
0.2
°
C. Thus a 1% resistor
of the proper value (4.369 K
Ω
for 45
°
C) can be used to set the span on the
measurement interface board. The release rate for the permeation tube is
then determined by weight loss in the actual oven being used.
You can use either of the two calibration methods presented here. One
method involves performing the “Setting Perm Oven Temperature”
procedure and then continuing with the “Setting Temperature with Water
Bath” procedure.
Alternatively, you can perform the “Setting Perm Oven Temperature”
procedure and then continue with the “Setting Temperature with Known
Resistance” procedure.
In either case, use the following procedure for setting the perm oven
temperature.
1.
Unplug POJ3 from the measurement interface board. Place a 4.369
K
Ω
resistor across pins 3 and 4 on the board.
2.
From the Permeation Oven menu in the Service menu, select
Cal
Oven Thermistor (
Figure 9–2
)
.
CAL OVEN THERM (RESISTOR):
CURRENTLY: 4369 Ohms
SET TO: 04369 Ohms
§¨
MOVE CURSOR
©ª
CHANGE VALUE
SAVE
RANGE AVG DIAGS ALARM
Figure 9–2.
Cal Oven Therm Resistor Screen
Permeation Tube
Oven Calibration
Setting Perm Oven
Temperature
Содержание 43i
Страница 4: ......
Страница 18: ...Contents Where to Get Help xviii Model 43i Trace Level Enhanced Instruction Manual Thermo Fisher Scientific ...
Страница 28: ...Introduction Specifications 1 6 Model 43i Trace Level Enhanced Instruction Manual Thermo Fisher Scientific ...
Страница 38: ...Installation Startup 2 10 Model 43i Trace Level Enhanced Instruction Manual Thermo Fisher Scientific ...
Страница 222: ...System Description I O Components 8 14 Model 43i Trace Level Enhanced Instruction Manual Thermo Fisher Scientific ...
Страница 238: ......
Страница 312: ......