C-Link Protocol Commands
Communications Configuration
B-34
Model 43
i
Trace Level-Enhanced Instruction Manual
Thermo Fisher Scientific
Send:
set
pmt
off
Receive:
set pmt off ok
save
set save params
This command stores all current parameters in FLASH memory. It is
important that each time instrument parameters are changed, that this
command be sent. If changes are not saved, they will be lost in the event of
a power failure. The example that follows saves the parameters to FLASH
memory.
Send:
set
save
params
Receive:
set save params ok
time
This command reports the current time (24-hour time). The example that
follows reports that the internal time is 2:15:30 pm.
Send:
time
Receive: time
14:15:30
set time
hh:mm:ss
hh
= hours
mm
= minutes
ss
= seconds
This command sets the internal clock (24-hour time). The example that
follows sets the internal time to 2:15 pm.
Note
If seconds are omitted, the seconds default to 00.
▲
Send:
set
time
14:15
Receive:
set time 14:15 ok
addr dns
This command reports the TCP/IP address for the domain name server.
Send:
addr
dns
Receive:
addr dns 192.168.1.1
set addr dns
address
This command sets the dns
address
, where
address
consists of four numbers
ranging from 0-255 inclusive, separated by “.”.
Communications
Configuration
Содержание 43i
Страница 4: ......
Страница 18: ...Contents Where to Get Help xviii Model 43i Trace Level Enhanced Instruction Manual Thermo Fisher Scientific ...
Страница 28: ...Introduction Specifications 1 6 Model 43i Trace Level Enhanced Instruction Manual Thermo Fisher Scientific ...
Страница 38: ...Installation Startup 2 10 Model 43i Trace Level Enhanced Instruction Manual Thermo Fisher Scientific ...
Страница 222: ...System Description I O Components 8 14 Model 43i Trace Level Enhanced Instruction Manual Thermo Fisher Scientific ...
Страница 238: ......
Страница 312: ......