Servicing
Flow Transducer Calibration
Thermo Fisher Scientific
Model 43i Trace Level-Enhanced Instruction Manual
7-31
2.
Disconnect the pump cable from AC PUMP connector on the
measurement interface board.
3.
From the Main Menu, press
to scroll to Service > press
>
to scroll to
Flow
Calibration
> and press
.
The Flow Sensor Cal screen appears.
Note
If Service Mode is not displayed, refer to “Accessing the Service
Mode” in this chapter, then return to the beginning of this step.
▲
4.
At the Flow Sensor Cal screen, press
to select
Zero
.
The Calibrate Flow Zero screen appears.
5.
Wait at least 10 seconds for the zero reading to stabilize, then press
to save the zero flow value.
6.
Reconnect the pump cable to the AC PUMP connector on the
measurement interface board.
7.
Connect a calibrated flow sensor at the SAMPLE bulkhead on the rear
panel.
8.
Press
to return to the Flow Sensor Cal screen.
9.
At the Flow Sensor Cal screen, press
to select
Span
.
The Calibrate Flow Span screen appears.
10.
Wait at least 10 seconds for the reading to stabilize, use
and
to enter the flow sensor reading, and press
to
save the value.
Содержание 43i
Страница 4: ......
Страница 18: ...Contents Where to Get Help xviii Model 43i Trace Level Enhanced Instruction Manual Thermo Fisher Scientific ...
Страница 28: ...Introduction Specifications 1 6 Model 43i Trace Level Enhanced Instruction Manual Thermo Fisher Scientific ...
Страница 38: ...Installation Startup 2 10 Model 43i Trace Level Enhanced Instruction Manual Thermo Fisher Scientific ...
Страница 222: ...System Description I O Components 8 14 Model 43i Trace Level Enhanced Instruction Manual Thermo Fisher Scientific ...
Страница 238: ......
Страница 312: ......