Operation
Alarms Menu
3-62
Model 43i Trace Level-Enhanced Instruction Manual
Thermo Fisher Scientific
The Lamp Voltage screen displays the current lamp voltage and sets the
minimum and maximum alarm limits. Acceptable alarm limits range from
500 to 1200 volts. If the lamp voltage goes beyond either the minimum or
maximum limit, an alarm is activated. The word “ALARM” appears in the
Run screen and in the Main Menu.
In the Main Menu, choose Alarms >
Lamp Voltage
.
LAMP VOLTAGE:
ACTUAL 1000 V
>MIN 750 V
MAX 1200 V
RANGE AVG DIAGS ALARM
The Minimum Lamp Voltage alarm limit screen is used to change the
minimum lamp voltage alarm limit. The minimum and maximum lamp
voltage screens function the same way.
In the Main Menu, choose Alarms > Lamp Voltage >
Min
or
Max
.
LAMP VOLTAGE:
ACTUAL MIN: 750 V
SET MIN TO: 700 V ?
©ª
INC/DEC
SAVE VALUE
RANGE AVG DIAGS ALARM
The Zero Span Check screen allows the user to view the status of the most
recent zero check and set the maximum zero check offset. The zero and
span check screens are visible only if the zero/span check option is enabled
and function the same way.
In the Main Menu, choose Alarms >
Zero
or
Span Check
.
Lamp Voltage
Min and Max Lamp
Voltage Limits
Zero and Span
Check
Содержание 43i
Страница 4: ......
Страница 18: ...Contents Where to Get Help xviii Model 43i Trace Level Enhanced Instruction Manual Thermo Fisher Scientific ...
Страница 28: ...Introduction Specifications 1 6 Model 43i Trace Level Enhanced Instruction Manual Thermo Fisher Scientific ...
Страница 38: ...Installation Startup 2 10 Model 43i Trace Level Enhanced Instruction Manual Thermo Fisher Scientific ...
Страница 222: ...System Description I O Components 8 14 Model 43i Trace Level Enhanced Instruction Manual Thermo Fisher Scientific ...
Страница 238: ......
Страница 312: ......