Operation
Alarms Menu
3-64
Model 43i Trace Level-Enhanced Instruction Manual
Thermo Fisher Scientific
goes beyond either the minimum or maximum limit, an alarm is activated.
The word “ALARM” appears in the Run screen and in the Main Menu.
In the Main Menu, choose Alarms >
SO
2
Concentration
.
SO2 CONCENTRATION:
ACTUAL 35.7
>MIN 100.0
MAX 100.0
MIN TRIGGER CEILING
RANGE AVG DIAGS ALARM
The Minimum SO
2
Concentration alarm limit screen is used to change the
minimum SO
2
concentration alarm limit. The minimum and maximum
SO
2
concentration alarm limit screens function the same way.
In the Main Menu, choose Alarms > SO
2
Concentration>
Min
or
Max
.
SO2 CONCENTRATION:
ACTUAL MIN: 100.0
SET MIN TO: 000400.
0
?
§¨
MOVE CURSOR
©ª
INC/DEC
SAVE VALUE
RANGE AVG DIAGS ALARM
The Minimum Trigger screen allows the user to view and set the SO
2
concentration alarm trigger type to either floor or ceiling. The minimum
alarm may be programmed as a floor trigger (alarm is triggered when the
concentration falls below the minimum value) or a ceiling trigger (alarm is
triggered when the concentration goes above the minimum value).
In the Main Menu, choose Alarms > SO
2
Concentration >
Min Trigger
.
MIN TRIG(CEILING/FLOOR):
ACTUAL TRIGGER: CEILING
SET TRIGGER TO: FLOOR ?
TOGGLE AND SAVE VALUE
RANGE AVG DIAGS ALARM
Min and Max SO
2
Concentration Limits
Min Trigger
Содержание 43i
Страница 4: ......
Страница 18: ...Contents Where to Get Help xviii Model 43i Trace Level Enhanced Instruction Manual Thermo Fisher Scientific ...
Страница 28: ...Introduction Specifications 1 6 Model 43i Trace Level Enhanced Instruction Manual Thermo Fisher Scientific ...
Страница 38: ...Installation Startup 2 10 Model 43i Trace Level Enhanced Instruction Manual Thermo Fisher Scientific ...
Страница 222: ...System Description I O Components 8 14 Model 43i Trace Level Enhanced Instruction Manual Thermo Fisher Scientific ...
Страница 238: ......
Страница 312: ......