Geysitech (Bayern-Hessen) Protocol
Geysitech Commands
Thermo Fisher Scientific
Model 43
i
Trace Level-Enhanced Instruction Manual
D-7
See
Table D–1
for operating status and
Table D–2
for error status for the
Model 43
i
Trace Level-Enhanced.
Table D–1.
Operating Status for Model 43
i
Trace Level-Enhanced
D7 D6 D5 D4 D3 D2 D1 D0
Æ
Bit
8
7
6
5
4
3
2
1
Æ
Hex-value
80 40 20 10 08 04 02 01
MSB
LSB
Operating status:
Service
Mode
(On)
0 0 0 0
0 0 0 1
Maintenance
(Local)
0 0 0 0
0 0 1 0
Zero
gas
(On)
0 0 0 0
0 1 0 0
Span
gas
(On)
0 0 0 0
1 0 0 0
Gas Unit Indication (ppb OR ppm) 0
0
0
1
0
0
0
0
Test
LED
(On)
0 0 1 0
0 0 0 0
Flash
Lamp
(Off)
0 1 0 0
0 0 0 0
Not
used
1 0 0 0
0 0 0 0
Table D–2.
Error Status for Model 43
i
Trace Level-Enhanced
D7 D6 D5 D4 D3 D2 D1 D0
Æ
Bit
8
7
6
5
4
3
2
1
Æ
Hex-value
80 40 20 10 08 04 02 01
MSB
LSB
Error status:
Internal
Temperature
Alarm
0 0 0 0
0 0 0 1
Reaction Chamber Temperature
Alarm
0 0 0 0
0 0 1 0
Lamp
Intensity
Alarm
0 0 0 0
0 1 0 0
Lamp
Voltage
Alarm
0 0 0 0
1 0 0 0
Pressure
Alarm
0 0 0 1
0 0 0 0
Sample
Flow
Alarm
0 0 1 0
0 0 0 0
Permeation Gas Temperature
Alarm
0 1 0 0
0 0 0 0
Not
used
1 0 0 0
0 0 0 0
Operating and Error
Status
Содержание 43i
Страница 4: ......
Страница 18: ...Contents Where to Get Help xviii Model 43i Trace Level Enhanced Instruction Manual Thermo Fisher Scientific ...
Страница 28: ...Introduction Specifications 1 6 Model 43i Trace Level Enhanced Instruction Manual Thermo Fisher Scientific ...
Страница 38: ...Installation Startup 2 10 Model 43i Trace Level Enhanced Instruction Manual Thermo Fisher Scientific ...
Страница 222: ...System Description I O Components 8 14 Model 43i Trace Level Enhanced Instruction Manual Thermo Fisher Scientific ...
Страница 238: ......
Страница 312: ......