Servicing
Flow Transducer Calibration
7-30
Model 43i Trace Level-Enhanced Instruction Manual
Thermo Fisher Scientific
Figure 7–14.
Replacing the Flow Transducer
Use the following procedure to calibrate the flow transducer.
Equipment Required:
Calibrated flow sensor
WARNING
The service procedures in this manual are restricted to qualified
representatives.
▲
If the equipment is operated in a manner not specified by the
manufacturer, the protection provided by the equipment may be
impaired.
▲
Equipment Damage
Some internal components can be damaged by small
amounts of static electricity. A properly ground antistatic wrist strap must
be worn while handling any internal component.
▲
1.
Remove the cover.
Flow Transducer
Calibration
Retaining Screws
Flow Transducer