Operation
Alarms Menu
3-58
Model 43i Trace Level-Enhanced Instruction Manual
Thermo Fisher Scientific
ALARMS:
ALARMS DETECTED 0
>INTERNAL TEMP OK
CHAMBER TEMP OK
PERM GAS TEMP OK
PRESSURE OK
SAMPLE FLOW OK
LAMP INTENSITY OK
LAMP VOLTAGE OK
ª
RANGE AVG DIAGS ALARM
ZERO CHECK OK
SPAN CHECK OK
ZERO AUTOCAL OK
SPAN AUTOCAL OK
SO2 CONCENTRATION OK
MOTHERBOARD STATUS OK
INTERFACE STATUS OK
I/O EXP STATUS OK
The internal temperature screen displays the current internal temperature
and sets the minimum and maximum alarm limits. Acceptable alarm limits
range from 8 to 47 C. If the internal temperature reading goes beyond
either the minimum or maximum alarm limit, an alarm is activated. The
word “ALARM” appears in the Run screen and in the Main Menu.
In the Main Menu, choose Alarms >
Internal Temp
.
INTERNAL TEMPERATURE:
ACTUAL 30.1
o
C
>MIN 8.0
o
C
MAX 47.0
o
C
RANGE AVG DIAGS ALARM
The Minimum Internal Temperature alarm limit screen is used to change
the minimum internal temperature alarm limit. The minimum and
maximum internal temperature screens function the same way.
In the Main Menu, choose Alarms > Internal Temp >
Min
or
Max
.
Internal
Temperature
Min and Max Internal
Temperature Limits
Содержание 43i
Страница 4: ......
Страница 18: ...Contents Where to Get Help xviii Model 43i Trace Level Enhanced Instruction Manual Thermo Fisher Scientific ...
Страница 28: ...Introduction Specifications 1 6 Model 43i Trace Level Enhanced Instruction Manual Thermo Fisher Scientific ...
Страница 38: ...Installation Startup 2 10 Model 43i Trace Level Enhanced Instruction Manual Thermo Fisher Scientific ...
Страница 222: ...System Description I O Components 8 14 Model 43i Trace Level Enhanced Instruction Manual Thermo Fisher Scientific ...
Страница 238: ......
Страница 312: ......