Servicing
I/O Expansion Board (Optional) Replacement
7-38
Model 43i Trace Level-Enhanced Instruction Manual
Thermo Fisher Scientific
Equipment Damage
Some internal components can be damaged by small
amounts of static electricity. A properly ground antistatic wrist strap must
be worn while handling any internal component.
▲
1.
Turn instrument OFF, unplug the power cord, and remove the cover.
2.
Unplug the I/O expansion board cable from the EXPANSION BD
connector on the motherboard.
3.
Remove the two standoffs holding the I/O expansion board connector
to the rear panel (
Figure 7–20
).
4.
Pop off the board from the mounting studs and remove the board.
5.
To install the I/O expansion board, follow previous steps in reverse.
6.
Calibrate the analog current outputs and analog voltage inputs per the
procedure earlier in this chapter.
Figure 7–19.
Replacing the I/O Expansion Board (Optional)
Motherboard
Digital Output Board
I/O Expansion Board (Opt)
Motherboard Support Bracket
Mounting Studs
Содержание 43i
Страница 4: ......
Страница 18: ...Contents Where to Get Help xviii Model 43i Trace Level Enhanced Instruction Manual Thermo Fisher Scientific ...
Страница 28: ...Introduction Specifications 1 6 Model 43i Trace Level Enhanced Instruction Manual Thermo Fisher Scientific ...
Страница 38: ...Installation Startup 2 10 Model 43i Trace Level Enhanced Instruction Manual Thermo Fisher Scientific ...
Страница 222: ...System Description I O Components 8 14 Model 43i Trace Level Enhanced Instruction Manual Thermo Fisher Scientific ...
Страница 238: ......
Страница 312: ......