goes to the Standby state. In the Standby state, the LC pumps and column oven are turned off and some mass
spectrometer voltages are turned off. The autosampler temperature control stays on to prevent sample degradation.
Only a user who has been assigned to the Administrator or Method Developer role can modify the length of time
the queue runs after the last acquisition has finished, before it puts the instrument in the Standby state.
1. Open the Configuration workspace.
2. Click
Queue
in the left panel.
3. Type the number of minutes that the system will remain in Ready state before moving to Standby state in the
Instrument idle time
field.
4. Type a number in the
Maximum number of acquired samples allowed
field.
5. Select the
If a sample is missing, proceed to the next sample
check box. The queue proceeds to
the next sample if a sample is missing.
6. Select the
If calibration fails, proceed to the next sample
check box. The queue proceeds to the next
sample if the calibration is failing.
7. Select the
Save calibration data (in the current project)
check box.
8. Click
Save
.
Select Regional Settings
This feature applies the region and language settings selected in Control Panel. Only a period “.” or comma “,”
can be used as a decimal separator. Digit grouping is not supported.
1. Open the Configuration workspace.
2. Click
General
in the left panel.
3. In the
General
panel, under
Regional Settings
, click
Apply
.
The regional settings that were set in the Windows operating system are applied to the software after the
computer is started again.
4. Click
Save
.
5. Start the computer again.
Manage the Compound Libraries
Import a LibraryView Package
1. Expand the
Compounds
list in the
Manage
pane.
2. Click
All Compounds
.
SCIEX X500 QTOF System
System User Guide
55 / 234
RUO-IDV-05-2334-E
Configure Access to the Software