WARNING! Radiation Hazard, Biohazard, or Toxic Chemical Hazard. Take care
to vent exhaust gases to a dedicated laboratory fume hood or exhaust system
and make sure that the ventilation tubing is secured with clamps. Make sure
that the laboratory has appropriate air exchange for the work performed.
Inspect the source exhaust drain bottle regularly, and empty it before it becomes full. Also inspect the bottle and
its fitting for leaks, and tighten connections or replace components, if required. Follow the steps in this procedure
to empty the bottle.
1. Remove the ion source. Refer to the ion source
Operator Guide
.
2. Loosen the clamps connecting the hoses to the cap of the source exhaust drain bottle.
3. Disconnect the hoses from the cap.
4. If applicable, lift the drain bottle out of the holder.
5. Remove the cap from the drain bottle.
6. Empty the drain bottle and then dispose of the waste according to laboratory procedures and local waste
regulations.
7. Install the cap on the bottle and then put the bottle in the holder.
8. Attach the hoses to the cap and then secure them tightly with clamps.