Table 6-19 Results Table Columns (continued)
LIS Supported
Description
Label
Y
(Optional) Shows the difference between the Found at Fragment
and the Fragment Mass, in ppm.
Fragment Mass Error
(ppm)
Y
(Optional) Shows the difference between the Found at Fragment
and the Fragment Mass, in mDa.
Fragment Mass Error
(mDa)
Y
Shows the detected peak height. If peak was not detected, then
this value is set to N/A.
Height
Y
For analytes using an internal standard, shows the ratio of the
Height to the IS Height. For internal standards, or for analytes
without an internal standard, this value is set to N/A.
Height Ratio
N
Shows the index of the row in the original, unsorted order. If the
table is sorted based on another column, then it can be returned
to the original order by sorting on this column.
This column is always visible in the Results Table. In the Column
Settings dialog, the check box is not available.
Index
Y
Shows the volume of the sample stored in the method and
injected by the autosampler.
Injection Volume
N
Shows the integration outliers.
Integration
Acceptance
Y
Shows the type of integration.
• Baseline: A standalone peak that was integrated in the usual
way.
• Valley: Indicates that there were two adjacent peaks and that
the signal did not return to the baseline value between them.
• N/A: Indicates that a peak was not detected.
• Manual: Indicates that the peak was manually integrated.
Integration Type
System User Guide
SCIEX X500 QTOF System
RUO-IDV-05-2334-E
160 / 234
Operating Instructions—Software