Reporter Templates
It is the responsibility of the user to validate the custom report template.
Some report templates use queries. Users can create queries using Microsoft Excel-based formulae to evaluate,
manipulate, and present the data from the Results Table in a report. The Metafield tag in the report template tells
the report the name of the query file that it should use. To use queries, the name of the query file must be specified
in the MetaField tag in the report template. Queries must also have the extension ".query" to be recognized as a
query. The queries must be stored in the Reporter folder where the report templates are stored.
We recommend that the user validate the generated results when a Reporter template is used, especially when
queries are used in a template. If any modifications are made to the report template after validation, then the
report template should be re-validated. Changes to the report template include any modification to reporter tags
or queries.
Table 6-27 Reporter Template Descriptions
Description
Template
A report showing, for each sample, a section including the File Information, Sample
Information, Analyte Results Table, and overlaid chromatograms of all of the analytes and
internal standard. The Analyte Results table is printed as shown in the Results Table. All
the qualitative confidence traffic lights are listed at the beginning of the table.
All Peaks Qual
A report showing the File Information for each analyte and an XIC table for each Blank,
each Standard, each QC, and 20% of the unknown samples. Unknown samples are selected
by the user-defined criteria in the report query.
Analyte 20 percent
A report showing, for each analyte, a section of the Samples Summary Table.
Analyte Summary
A report showing the File Information, Statistics Table (standards), and Calibration Curve
for analytes, one page per analyte.
Calibration Curves
A report showing, for each analyte, a section including the File Information and a metric
plot of the analyte peak area.
Metric Plot
A report showing, for each selected sample, a section including the File Information,
Sample Information, and Analyte Results Table for the selected analytes. The Analyte
Results table is printed as shown in the Results Table. All the qualitative confidence traffic
lights are listed at the beginning of the table.
Per Sample
Quant-Qual
A report showing, for each selected sample, a section including the File Information,
Sample Information, and Analyte Results Table for the selected analytes. The Analyte
Results table is printed as shown in the Results Table. All the qualitative confidence traffic
lights are listed at the beginning of the table.
Per Sample
Quant-Qual Visible
Rows Using Visible
Analyte