Prepare the Mass Spectrometer
WARNING! Hot Surface Hazard. Let the ion source and vacuum interface cool for at
least 30 minutes before starting any maintenance procedures. Some surfaces of the
ion source and vacuum interface become hot during operation.
CAUTION: Potential System Damage. Do not drop anything into the source drain when the
ion source is removed.
Figure 7-2 Source Drain on the Vacuum Interface
1. Deactivate the devices. Refer to
2. Remove the ion source. Refer to the ion source
Operator Guide
.
When the ion source is not in use, store it to protect it from damage and to maintain operating integrity.
SCIEX X500 QTOF System
System User Guide
205 / 234
RUO-IDV-05-2334-E
Service and Maintenance Information