1. Open the MS Tune workspace.
2. Select
Positive Quick Status Check
or
Negative Quick Status Check
from the
Tuning Procedures
list.
3. Double-click the CDS power icon ( ) on the status panel to turn on the CDS flow.
4. Select the correct channel, 1 or 2, click
Start
, and then close the window.
5. Wait at least 1 minute for the CDS to equilibrate.
6. Click
Next
.
7. Follow the on-screen instructions for each step.
8. (Optional ) Review the report to verify the results of each step.
9. (Optional) Save the report.
10. Click
Save Tuning Settings
if the results are satisfactory. If the results are not satisfactory, then do one of
the following:
• Repeat the steps.
• Run the TOF MS tuning procedure. Refer to
• Discard the results by closing the
MS Tune
workspace.
• Restore the previous settings by selecting the appropriate backup file from the
Restore Instrument
Data
menu.
Optimize the Detector
When the system sensitivity is low, use this procedure to verify that the detector voltage is optimized. During the
procedure, the software can adjust the detector voltage to provide the optimum sensitivity. When the optimization
is completed, the user can save the optimized value or discard the changes.
We recommend optimizing the detector every 6 months, or when the signal diminishes or is lost after cleaning.
1. Open the MS Tune workspace.
2. Select
Detector Optimization
from the
Tuning Procedures
list.
3. Make sure that the spray is stable.
4. Click
Next
.
5. Follow the on-screen instructions.
6. Click
Next
.
The optimization report is shown.
7. (Optional) Save the report.
SCIEX X500 QTOF System
System User Guide
193 / 234
RUO-IDV-05-2334-E
Operating Instructions—Software