Table 6-27 Reporter Template Descriptions (continued)
Description
Template
A report showing, for each analyte, a section including the File Information, Results Table,
Calibration Curves, and chromatograms including the internal standard and each analyte.
This template is suitable for a Results Table with a group defined in it.
Per Analyte
Quant-Qual
A report showing, for each sample, a section including the File Information, Sample
Information, Analyte Results Table, Calibration Curves for each analyte, and chromatograms
including the internal standard and each analyte. This template is suitable for a Results
Table with a group defined in it.
Per Sample Report
A report showing, for each selected sample, a section including the File Information;
Sample Information; Analyte Results Table for the selected analytes; overlaid
chromatograms of all of the analytes, internal standard, and the XIC; the Acquired/
Theoretical MS spectra; and the Acquired/Library MS/MS spectra for each selected analyte.
The Analyte Results table is printed as shown in the Results Table. All the qualitative
confidence traffic lights are listed at the beginning of the table.
Positive Hits Qual
A report in .csv format showing, for each sample, a section including the File Information,
Sample Information, and Analyte Results Table.
Qual CSV report
A report showing, for each sample, a section of Analytes Summary Table. This report
template is suitable for a Results Table with groups.
Sample Summary
MS Tune
A .dat file is created by the software when the instrument data is saved. Use this file to restore earlier parameter
states. The .dat backup file is named using the time that the file was created, not the time that the file was backed
up.
Note:
Instrument optimization should be performed using the ESI probe only.
Each time the user loads the MS Tune procedure, all of the mass spectrometer parameters are backed up.
Perform a Quick Status Check
Use this procedure to calibrate the system and to quickly verify the resolution in TOF MS and MS/MS modes. If
the channel alignment mass accuracy does not meet the specification, then the user can repeat the steps and
calibrate the system. If the resolution does not meet the specification, then the user can perform the TOF Tuning
procedure to optimize the system.
Tip!
Users can assess the this procedure by clicking
MS Check
on the status panel.
System User Guide
SCIEX X500 QTOF System
RUO-IDV-05-2334-E
192 / 234
Operating Instructions—Software