1. Turn the bottle counter-clockwise to remove it from the CDS.
2. Turn the new bottle clockwise to install it.
Start the CDS
Use the direct control function to start the CDS manually when flushing the CDS or when introducing solutions
during tuning.
1. On the status panel, click the
Direct CDS Control
icon ( ).
The Device Control dialog opens.
Figure 3-6 Device Control (CDS)
2. Click
Start
.
Stop the CDS
1. On the status panel, click the
Direct CDS Control
( ) icon.
2. Click
Stop
.
Flush the CDS
WARNING! Toxic Chemical Hazard. Refer to the chemical product Safety Data Sheets
and follow all of the recommended safety procedures when handling, storing, and
disposing of chemicals. For health and safety precautions, refer to
.